Temperature sensors TVO (TBO) |
|
- Temperature calibrate coal resistors
Temperature calibrated coal resistors TVO are proposed as temperature sensors to be used in various types of cryostats and variable temperature cryostats-inserts. The calibration is performed at 20 temperature points by the use of the Rh secondary temperature standard. The polynomial approximation of experimental points ensures small temperature step and high accuracy in temperature measurements. The highly reliable performance of the coal resistors allows hundred thousands thermal cycles in the temperature region 1.8 – 300 K without any changes in mechanical properties and temperature calibration.
Fig.1. Temperature sensor TVO
3-points calibration
Application:
Safety control for superconducting solenoids
-
MR imaging unit
-
ESR spectrometer
-
NMR spectrometer
Calibration points:
20-points calibration
Fig.2. Temperature-resistance dependence for the sensor TVO in T = 1.8 – 300 K
Parameters:
Calibration:
-
20 calibration points
-
Polynomial fit coefficients
-
Fitting points interval ΔT = 0.01 K
-
Number of the calibration points can be increased
Technical parameters
Resistance at room temperature, Ohm
|
910
|
Resistance spread at room temperature
|
5%
|
Resistance shift in the magnetic field 8 T at 4.2 K
|
4%
|
Repeatability of measurements after 1 year at Т = 4.2 K, K
|
0.01
|
Sensor dimensions, mm
|
1 х 2.5 х 8
|
Whole length of resistors including current leads and potential contacts soldering, mm
|
60
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Postal address: |
RTI Ltd., Institutskaya str. b.2, 142432 Chernogolovka, Moscow region, Russian Federation |
Phone: |
+7 800 100-29-70 |
|
|
|
|
|
|
| |